High Frequency Digital Radiography
BT-PLD3600High Frequency Digital Radiography System
This machine is applied to take digital radiography on every part of human body, such as head, limbs, chest, limbus and abdomen and etc.
Item |
Content |
Technical Parameter |
|||
Power Supply |
Voltage |
220V |
|||
Frequency |
50Hz |
||||
Capacity |
≥40kVA |
||||
internal resistor |
≤0.15Ω |
||||
Generator |
|
Power Output |
32KW |
||
Inverter Frequency |
30 KHz |
||||
photography |
Tube Voltage |
40kV—150kV |
|||
Tube Current |
10mA—320mA |
||||
Exposure time |
1.0ms—6300ms |
||||
Digital Controlled X-ray Tube |
Tube Focus |
1.2mm /0.6mm(Large/Small Focus) |
|||
Input Power |
Large/Small focus:40kW/20kW |
||||
Thermal capacity |
110KJ |
||||
Rotary anode speed |
2800rpm |
||||
Radiography Table |
Table longitudinal movement |
≥900mm |
|||
Table transverse movement |
≥220mm |
||||
Pillar movementalong table |
≥1300mm |
||||
Detectingholdermovement |
≥500mm |
||||
Tube assembly up-down movement |
500mm-1280mm |
||||
Collimator |
Manual Multi-leaf |
||||
Table use fixed grid |
Grid density:103L/INCH Grid ratio:10:1 SID:120cm Stationary type:15″×18″ |
||||
Bucky Stand |
Radio-graphic device movement along pillar |
≥1300mm |
|||
Cassette SID |
450-1780mm |
||||
Fixed grid |
Grid density103L/INCH Grid ratio:10:1 SID:180cm Size:15″×18″ |
||||
Flat Panel Detector |
Vision size |
358mm(H)×430 mm(V) |
|||
Pixel Matrix |
2560(H)×3072(V) |
||||
Pixel Pitch |
139 μm |
||||
Imaging time |
less than 10s |
||||
Limiting resolution |
3.5 lp/mm typ |
||||
A / D transform |
14 bit |
||||
Energy range |
40 - 150 kVp |
||||
Max input dose (low transmission gain) |
4 mR / frame |
||||
Date out |
A 16-bit digital output Ethernet(1000BASE - T) |
||||
Command control |
Ethernet(1000BASE - T) |
||||
Power input |
DC 24V 4.8A |
||||
Work Station |
CPU System |
Brand:DELL OPTIPLEX 7010commercial use Processor:Core™ i5-3570processor RAM:4GB DDR3 Hardware:500GB SATA (7200 rpm) CD-driver:suitMT’s16X variable-speed DVD + / - RW With functions of double written Network Card:Broadcom NetXtreme 10/100/1000 PCIe Gigabit LAN GNTB-A High display card:1GB AMD RADEON HD 7570,FH,with DVI-VGA Adapter Slop:PCIEseriel port and parallel port,all height,MT |
|||
Color LCD Monitor |
19inches1M LEDbacking light LCD display revolution1280x1024,5:4screen dot pitch: 0.294mm contrast ratio:1000:1 brightness:250cd/㎡ Grey feedback time: 5ms visual Angle:160/170° |
||||
Monochrome Monitor(Optional) |
1280*1024,19inches brightness1000cd/m2 Contrast Ratio900:1 |
||||
Work Station Software |
Basic operation: Change control console password, edit ID,acquisite images. |
||||
Additional operation: Add new check, edit present checking info, add new position, change image aquisition order, multi-checking agreement chance, manual adjustment of exposure parameters, automatic exposure control mode, focus choice, patient body-type choice, tube capacity check, ESA curve choice, image cutting, note added (sent to DICOM workstation), mark on images, rotate or overturn, full-size image observation, check patient info and dose info, accept or refuse images. |
|||||
Image management: Change order, patient basic info editing, inquiry history images, resend history images, re-print history images, check images mark info, review history images, manage refused images, space reclaimed, image protection, and manual image deletion etc. |
|||||
System management: ID edition, change ID password, ED refrigeration set,statisticsinfo checking, detector calibration, equipment control, output order management, image measurement. |